User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: LFE Series 40:Digital Panel Meter DATE: 02-Oct-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 14 NUMBER OF LINES: 40 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- F W 1.002 ASK+ X 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP [32] Connect the 5700A and the UUT as follows: 1.004 DISP [32][27][91]1m 5700A TO LFE 40 1.004 DISP [32] OUTPUT V Ohms A HI ......... PIN M 1.004 DISP [32] OUTPUT V Ohms A LO ......... COM N 1.005 DISP Connect the UUT to 115 VAC. 1.006 HEAD {* DISPLAY TEST *} 1.007 MESS Adjust the 5700A output for a UUT display of 188.8. 1.008 5700 188.8V N 2W 1.009 MESS 1.010 5700 * S 1.011 EVAL {Did all display segments illuminate?} 2.001 HEAD {* DC VOLTS FULL SCALE AND LINEARITY TEST *} 2.002 5700 199.00V 0.05% 0.1U 2W 3.001 5700 150.00V 0.05% 0.1U 2W 4.001 5700 99.90V 0.05% 0.1U 2W 5.001 5700 88.80V 0.05% 0.1U 2W 6.001 5700 77.70V 0.05% 0.1U 2W 7.001 5700 66.60V 0.05% 0.1U 2W 8.001 5700 55.50V 0.05% 0.1U 2W 9.001 5700 44.40V 0.05% 0.1U 2W 10.001 5700 33.30V 0.05% 0.1U 2W 11.001 5700 22.20V 0.05% 0.1U 2W 12.001 5700 11.10V 0.05% 0.1U 2W 13.001 5700 0.10V 0.05% 0.1U 2W 14.001 END